EX2000 ENMET Corporation
5
3.1.1 Alarms
Audio and visual alarms will activate in pulsed or continuous, to indicate the type of alarm being triggered. The display will
indicate the threshold or code of the alarm. To clear a threshold alarm after leaving the dangerous area, press and release the
ALARM
switch.
N
OTE
: If the alarm threshold is still exceeded: The visual alarm will continue to flash and the threshold pictogram will remain
indicated.
Fault alarms are not able to be cleared, see fault alarms table for possible remedies.
Threshold Alarms: Pulsed audio alarm can be cleared by pressing the
ALARM
switch
L
Alarm Type Alarm Indication Parameters
Threshold 1
Display:
XXX
%
Audio : 1 beep every 2 seconds
Adjustable from 0 – 60% LEL
instantaneous alarm
(adjustable)
Threshold 2
Display:
XXX
%
Audio: 1 beep per second
Fixed at 60 % LEL
(factory set, not adjustable)
Over-range SUP; continuous audio alarm See fault alarms table
Fault Alarms: Continuous audio alarm can not be cleared
Display Possible Cause Possible Remedy
Display: BAT
Batteries discharged:
The instrument can no longer operate
Turn instrument off and recharge the
batteries
Display: SUP Measurement range is exceeded Leave contaminated area and allow
sensor to stabilize in clean air.
1DEF The cell zero has drifted out of spec Calibrate
2DEF The cell is too sensitive Calibrate
4DEF (on small digits) Expired Cell (sensitivity too low) Replace cell and calibrate
8DEF
The EE
PROM
of the instrument is not
initialized
Return to ENMET for initialization
512DEF Problem in the calculations, capacity
exceeded
Return to ENMET
16DEF Measurement to negative Calibrate
May need to replace cell and calibrate
64DEF Temperature sensor is damaged Return to ENMET
Er45 or Er49 Communication error between the
microprocessor and the cell EE
PROM
• Disconnect the cell and check the
connection.
• Replace cell if message persists
• Return to ENMET
N
OTE
: If several faults are simultaneously present, the display corresponds to the sum of faults.
Example: The messages Er 35 or Er 39 show the communication errors between the microcontroller of the apparatus and
the EE
PROM
memory of the device.