ISA-40M Oxygen Monitor ENMET Corporation
3
3.0 Installation3.0 Installation
WARNING: This equipment is not for use in hazardous combustible atmospheres as defined by the national
electrical code. Use in such atmospheres may result in property damage, injury or death
Figure 2: Mounting Dimensions
3.1 Control Unit Mounting and Power Hook-up
Use an appropriate fitting for conduit when supplying power to the unit.
1. Choose a suitable location for mounting the electronics control unit. See Figure 2 for mounting dimensions.
NOTE: All dimensions are given in inches.
2. Apply 110 VAC and/or 12 VDC power to the appropriate terminals on TB5 and TB3. Refer to the Wiring
Information table. Both 110 VAC and 12 VDC power can be applied together; the 12 VDC is standby power
source; DC current will flow only when the AC power is interrupted.
Table 1: Power Supply Wiring
Power supply ISA-40M Control Unit connection
110 VAC HOT TB5 HOT
110 VAC NEUTRAL TB5 NEUT
AC ground TB5 GND
12 VDC TB3-3 + VDC
DC ground TB3-1 or TB3-2
4 places
Fuse
Circuit Breakers
Transformer
Protective
Standoff cover
TB3
TB5