ENMET Corporation QUADRANT
8
recharge the battery
HGAM
alternated with
>100 GAS
overrange of combustible gas
-Unacknowledgeable visual and
audible alarms in combustible
gases
-Acknowledgeable visual and
audible alarms in oxygen and
toxic gases
leave dangerous area and turn off
instrument, it may then be
restarted
GAS faul
The sensor is defective or absent
-Unaknowledgeable visual and
audible alarms
turn off instrument and install or
change associated sensor
disable channel
Calibration is unsuccessful
Measurement is too negative
perform new calibration: the
sensor must be changed if fault
persists
GAS used
Sensor is worn out
-Unaknowledgeable visual and
audible alarms
turn off instrument and change
sensor
disable channel
GAS CAL
Auto -zero is unsuccessful
-Unaknowledgeable visual and
audible alarms
repeat Auto-zero
change sensor or cell
FAULT Lithium battery is worn out
Memory loss
-Unaknowledgeable visual and
audible alarms
change the lithium battery
the instrument must be returned
to ENMET if fault persists