Appendix A: Calibration Data Tables
Appendix A: Calibration Data TablesAppendix A: Calibration Data Tables
Appendix A: Calibration Data Tables
Table 1: Gas Ranges, Alarm Points and Sensor Life
Gas Range
Alarm
lower limit
Alarm
upper limit
Factory Alarm
Set Point
Expected Sensor
Lifetime
Ammonia
0-100ppm 15ppm 50ppm 25ppm 1 year
Arsine
0-0.99ppm 0.03ppm 0.10ppm 0.05ppm 1-1.5 years
Carbon Monoxide
0-1000ppm 5ppm 200ppm 35ppm 2-3 years
Chlorine
0-10ppm 0.5ppm 5.0ppm 1.0ppm 1-2 years
Fluorine
0-10ppm 1ppm 5ppm 1ppm 2-3 years
Hydrogen
0-1000ppm 200ppm 750ppm 200ppm 1-2 years
Hydrogen Chloride x
xx
x
0-20ppm 3.0ppm 10.0ppm 5.0ppm 1-2 years
Hydrogen Cyanide
0-20ppm 3.0ppm 10ppm 4.7ppm 2-3 years
Hydrogen Fluoride
0-10ppm 2.0ppm 6.0ppm 3.0ppm 1 year
Hydrogen Sulfide
0-200ppm 5.0ppm 25ppm 10ppm 2-3 years
Nitric Oxide x
xx
x
0-100ppm 15ppm 50ppm 25ppm 1-2 years
Nitrogen Dioxide
0-10ppm 1.0ppm 5.0ppm 3.0ppm 1-2 years
Oxygen
0-25% 16% 23.5% 19.5% & 23.5% 1.5-2 years
Ozone
0-0.99ppm 0.05ppm 0.2ppm 0.1ppm 1-1.5 years
Phosgene
0-0.99ppm 0.05ppm 0.20ppm 0.1ppm 1 years
Phosphine
0-0.99ppm 0.15ppm 0.30ppm 0.50ppm 1-1.5 years
Silane
0-30ppm 3ppm 5ppm 5ppm 1-1.5 years
Sulfur Dioxide
0-20ppm 2.0ppm 10ppm 5ppm 1-2 years
N
OTE
: All specifications stated in this manual may change without notice.
x
xx
x Biased Sensors: Instruments using biased sensor, stabilization time is extended to 4 minutes. Complete stabilization
may take as long as 1 hour. When instrument is turned on it may display XXXX
+
. If instrument
has been off for an extended amount of time it may be necessary to “cycle” (turn on and off) the
instrument several times.
See Appendix C for Hydrazine