5.2.2 Gas Span
It is recommended that the Zero Function be performed first.
Enter the maintenance menu. See
Figure 5,
EX-5120
Maintenance Menu flow chart.
1.
Tap the
M
ENU
switch once to show Span on the display.
2.
Tap the
S
ELECT
switch to perform a Span procedure. The display will alternate between the calibration gas concentration
and a signal level.
3.
Attach the associated calibration gas cylinder to the regulator and calibration cover. See to
Figure
4
.
4.
Open the valve to apply the calibration gas to the sensor.
5.
Watch for the signal level to stabilize.
6.
Once the signal level has stabilized, the
EX-5120
will automatically lock in the calibration data and:
If the Span is successful, Cal OK appears on the display momentarily, then advances to Alarm 1. Remove calibration gas.
To exit maintenance menu tap the
M
ENU
switch until Exit appears, then tap the
S
ELECT
switch.
If the sensor is outside of acceptable parameters, Bad Span is displayed momentarily, then returns to Span.
Remove calibration gas. Tap the
M
ENU
switch until Exit appears, then tap the
S
ELECT
switch. Check span gas and
repeat calibration.
N
OTE
:
Some software revisions require the
S
ELECT
switch be tapped to accept the signal.
7.
Calibration is complete.
5.2.3 Exit Maintenance Menu
Exit maintenance, by tapping on the
M
ENU
switch until Exit appears on the display. Tap the
S
ELECT
switch to return to the
instrument Normal Gas Display.
Normal Display Mode
N
OTE
:
You can change the Calibration Gas Level. H
OLD
the magnet over the S
ELECT
switch for 2 – 4 seconds
The M
ENU
switch changes digit indicated by underscore cursor
The S
ELECT
switch locks underscored digit and moves to next digit
5%
M
ENU
H
OLD
the magnet over the
M
ENU
switch for 2 – 4 seconds
to enter the Maintenance Menus
Exit
S
ELECT
Tap the
S
ELECT
switch to return to the Normal
Display Mode.
See Section 5.2.3
M
ENU
Tap the magnet over the M
ENU
switch to cycle through Maintenance Menus
OR
Zero
Cal OK
If
the Zero signal is within Preset Specs the EX-5120 will display
Cal OK, See Section 5.2.1
If the Zero signal is not within Preset Specs the EX-5120 will
display Bad ZERO
N
OTE
:
Some software revisions require the
S
ELECT
switch be
tapped to accept the signal.
S
ELECT
M
ENU
Bad ZERO
PV: 0
S
ELECT
Tap the S
ELECT
switch to initiate Zero adjustment