EX-5185-PID
ENMET Corporation
12
Normal Display Mode
5ppm
M
ENU
OLD
the magnet over the
ENU
switch for 2 – 4 seconds
to enter the Maintenance Menus
Exit
S
ELECT
Tap the
ELECT
switch to return to the Normal
Display Mode.
See Section 5.2.3
M
ENU
Tap the magnet over the M
ENU
switch to cycle through Maintenance Menus
OR
Zero
the Zero signal is within Preset Specs the EX-5185-PID will
display Cal OK, See Section 5.2.1
If the Zero signal is not within Preset Specs the EX-5185-PID will
display Bad ZERO
N
OTE
:
Some software revisions require the
S
ELECT
switch be
tapped to accept the signal.
S
ELECT
M
ENU
Bad ZERO
S
ELECT
Tap the S
ELECT
switch to initiate Zero adjustment
Calibration Gas until signal value becomes
stable (about 1 to 4 minutes)
See
Figure
4
If cal is good display will indicate OK or Same
If cal is not within preset “range” display will
indicate Bad Sens
N
OTE
:
Some software revisions require the
S
ELECT
switch
be tapped to accept the signal.
M
ENU
PV: 0
20
Cal OK
Same mV
Bad Sens
Span
S
ELECT
S
ELECT
OR
OR
N
OTE
: You can change the Calibration Gas Level.
H
OLD
the magnet over the S
ELECT
switch
The M
ENU
switch changes digit indicated by underscore cursor
The S
ELECT
switch locks underscored digit and moves to next digit
See Section 5.2.2
S
ELECT
Menu
To change Alarm set points:
Tap Menu switch until Alarm to be changed is displayed
Tap Select switch to display the set point
The M
ENU
switch: changes digit indicated by underscore
cursor
The S
ELECT
switch: locks in the underscored digit and
moves to next digit
If change is not within range display returns to first digit
If change is within range display moves to Set Time Delay
Use M
ENU
and S
ELECT
switches as above to change time
delay. Between 0 and 5 seconds is allowed
If change is within range display moves to next menu
Λ - Indicates increasing alarm
V - Indicates decreasing alarm
Alarm3
S
ELECT
Λ 50
M
ENU
SetTDsec
0
Alarm1
M
ENU
S
ELECT
Λ
M
ENU
S
ELECT
Λ 20
Figure 5: Continued on next page